FREESTYLE PORTABLE OXYGEN CONCENTRATOR
17
Alarm/Light Indicators
When the FreeStyle unit senses inhalation, a pulse of oxygen is sent through the nasal cannula. The green light above the
selected pulse flow setting displays intermittently each time a breath is detected.
Additionally, when the unit is operating on and simultaneously being charged through the AC or DC power supply, the
FreeStyle unit's battery gauge/indicator(s) display the charge level of the battery and remain on for approximately one hour
after reaching a full charge.
An audible alarm sounds if FreeStyle has a low battery, if the cannula is disconnected, or if performance of the unit is outside
specifications. The light and audible alarm conditions are explained in detail below and summarized on the chart later in this
section of the manual.
Start-Up
A brief alarm sounds at start-up. FreeStyle begins to operate when the alarm stops.
Low battery
As the battery power approaches a low level, a brief alarm sounds intermittently, and the yellow 25% BATT
light also illuminates intermittently. When this occurs, connect FreeStyle into a DC power outlet or an AC
power outlet, or change to another source of oxygen within two minutes. When FreeStyle is connected to DC
power outlet or to an AC power outlet, the unit operates while recharging FreeStyle’s battery simultaneously.
The level of battery charge is indicated by the battery gauge/indicator(s).