  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 37
MN123-1 rev E 11/12
34
Problem
Probable Cause
Solution
FreeStyle does not operate when a pulse flow
selection button is pressed.
Battery is discharged.
Malfunction.
Power the unit through the DC outlet, or an AC
outlet.
Contact your Equipment Provider, and change to
another source of oxygen as necessary.
A continuous alarm sounds and the (red) alarm
light illuminates continuously.
FreeStyle has not detected a breath for 15
minutes.
Check the cannula connection.
Ensure that cannula tubing is not kinked.
Make sure that the cannula is positioned properly
and that you are breathing through your nose. For
mouth-breathing your clinician may recommend a
chin strap.
Intermittent alarm condition, and the (yellow)
BATT light illuminates intermittently.
Battery requires charging.
Connect to a DC or an AC outlet within 2 minutes,
or connect to the optional AirBelt.
Do not attempt any maintenance other than the possible solutions listed below.